VISIONC600 - VisionR700 & VisionR800 keyboard can't control the screen.

This problem can only occur if the screen VISIONC600 is configured with the VisionR using network method.
When the screen VISIONC600 is not available the chart library is disabled.
The letter (S) is not showing.
Possible reasons
1. Screen VISIONC600 is turned off or no power is detected.
2. Screen VISIONC600 is disconnected from Network OR Ethernet Cable is disconnected.
3. Screen VISIONC600 IP changed and the VisionR can't find it.
4. VisionR can't find hostname (Address) due to is disconnected from Network OR Ethernet Cable is disconnected.
Solution
Related Articles
CSPOLA600 - Why the screen display white?
Make sure you have a USB 3.0 THUMB DRIVE. Make sure to use software V1.0.0.1 to update the CSPOLA, to be used with Vision R 800 V1.2.0.1 1. Use the power switch to turn off the unit. 2. Proceed to download file V1.0.0.1.zip and then place it in the ...
CSPOLA600 - Why is the screen black?
1) Check if the power source is working. Voltage output 12V. 2) Check if the black screen appears after turning on the power. 3) It is possible that the PN CSP622 Converter board needs to be replaced
CSPOLA600 - Product will not respond to remote control
Possible Causes IR detector is dirty. Batteries in remote are dead Suggested Solutions Clean the IR detector. Recharge the remote control using the USB cable. Replace the battery
CS550 - Product will not respond to remote control
Possible Causes IR detector is dirty. Batteries in remote are dead Suggested Solutions Clean the IR detector. Recharge the remote control using the USB cable. Replace the battery
CSPOLA600 - Don’t want the slideshow during sleep mode
Possible Causes Needs to be turned off Suggested Solutions In system setup, change the “Slideshow” to “OFF”. Refer to user manual for instructions.