SL650+ - How to use Dry Eye - Instructions

SL650+ - How to use Dry Eye - Instructions

NIBUT

1) Set the attachment to Placido Ring.
2) Set the Mags to 10X.
3) Ask the patient to blink once after starting the video.
4) Perform a 20 Seconds Video.



Tear Height

1) Set the attachment to Placido Ring.
2) Set the Mags to 10X.
3) Ask the patient to blink once.
4) Take picture. Make sure to display the tear height in the window.



Lipid Layer

1) Set the attachment to White Ring.
2) Set the Mags to 10X.
3) Ask the patient to blink once every 2 seconds, after starting the video.
4) Perform a 10 Seconds Video.



Mebomian Glands

1) Light Slit or Attachment is not needed.
2) Set the Mags to 6X.
3) Ask the patient to stop blinking
4) Take picture of Upper meibomian glands.
5) Take picture of Lower meibomian glands.


Eyelid Edge

1) Light Slit or Attachment is not needed.
2) Use the background illumination only.
3) Set the Mags to 10X.
4) Ask the patient to stop blinking
5) Take picture of Upper eyelid.
6) Take picture of Lower eyelid.


Eye Redness

1) Light Slit or Attachment is not needed.
2) Use the background illumination only.
3) Set the Mags to 6X.
4) Ask the patient to stop blinking
5) Take picture of whole eye.


Ocular Surface Staining

1) Attachment is not needed.
2) Use diffuser lens and maximum brightness. 14 mm light spot.
3) Set the Mags to 6X.
4) Set blue & yellow filter.
4) Ask the patient to stop blinking
5) Take picture of whole eye.